Constant temperature and humidity chamber humidity can not rise to the set value how to do?

In the process of using a constant temperature and humidity chamber in a laboratory, there may be some malfunctioning problems. For example, the humidity does not rise to the set value.

When the humidity does not rise to the set value, you can check in several ways.

First, you can check the constant temperature and humidity oven water level height, its water level should be in the middle of the heating tube on the upper; if in the heating tube on the lower side, may lead to slow humidification, and can not add up to the situation. This time the solution is relatively simple, that is, to add water in a timely manner.

The second is to check the humidification electric heating tube, power: 300 W. Electric heating tube burned, you can contact the manufacturer of the product.

The third, is to check the instrument, is not damaged, resulting in data failure.

Fourth, at the same time, it is necessary to note that the users do not frequently change the set value, so as to avoid the frequent start of the compressor caused by overload, affecting the service life of the equipment.