What is the cause of the water running down the bottom of the biochemical incubator?

Biochemical incubator in the work project, if the case of flowing water, it is necessary to analyze what went wrong.

Biochemical incubator temperature range 0 ~ 65 degrees, generally we use the temperature in the laboratory and then between 25-37 degrees, the ambient temperature of 30 degrees is if the use of temperature 25 degrees constant temperature, the incubator to cool down, cooling is done through the refrigeration unit.

Refrigeration produces temperature difference will have condensation, ready-made water. The biochemical incubator has a water catchment port, water through the water catchment port discharged to the outside of the incubator, will not flow to the bottom of the biochemical incubator. What is the reason when water is found at the bottom of the biochemical incubator? It is the water produced by the refrigeration unit on the evaporator in the incubator flows along the copper tube to the bottom of the incubator, and it is the copper tube lead outlet in the incubator that is not closed properly.