The difference between the three drying ovens and how to use them

 1. The difference between the control panel of the drying oven GX, DHG and KH series

Constant temperature drying oven GX, DHG, KH series, divided into LCD screen and digital display screen, using microcomputer intelligent PID technology, the LCD screen temperature control accuracy is ±0.1 degrees, with RS485 microcomputer interface function, can realize network management and monitoring, etc. The controllable accuracy of the digital display meter is ±0.5 degrees. In terms of accuracy and function, the LCD meter is obviously better than the digital display meter.

2. Then about the setting method of the temperature and timing of the drying oven instrument:

Press the Est setting key first, the SP setting temperature appears, click to save, and the SE setting time appears, click to save, and the time is set to 0, which means that it is irregular.

3. Our oven has an over-temperature alarm function. If the temperature exceeds the temperature, the instrument will buzz and alarm. If you need sound and light alarms, you can install a sound and light alarm system. In order to avoid some customers dislike the big impact of sound, we do not have audible and visual alarms at the factory, only the buzzer and over-temperature alarm that comes with the instrument.

4. The factory voltage of the product is set at 220V, 50Hz, and can be changed to 110V, 60Hz according to customer requirements, which is suitable for the voltage requirements of many countries.

 drying oven:http://en.kentonchina.com/list/drying-oven.html

评论

此博客中的热门博文

What incubator should I use for fungi?

Laminar flow Cabinet architecture introduction

The Control system is the core of the constant temperature and humidity chamber